Catalog Number A105100000000
Device Problems Detachment of Device or Device Component (2907); Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a x-smart iq contra angle won't hold files; no injury resulted.
 
Manufacturer Narrative
X-smart iq ca head gear assy.Various mechanical problem.Ball bearing defect.Replaced 1 x iq contra-angle a105100000000.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-SMART IQ CONTRA ANGLE
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
TULSA DENTAL PRODUCTS LLC
608 rolling hills drive
johnson city TN 37604
MDR Report Key10519128
MDR Text Key215579035
Report Number8031010-2020-00170
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
PMA/PMN Number
K161213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA105100000000
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/05/2020
Initial Date FDA Received09/10/2020
Supplement Dates Manufacturer Received08/05/2020
Supplement Dates FDA Received10/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-