The device evaluation is anticipated.However, the complaint cannot not be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation results when received.Lot number was not provided, therefore review of the manufacturing records could not be completed.
|
It was reported that the swan catheter and pressure transducer gave inaccurate pressures.Troubleshooting included switching out mac lab, swan/transducer cable, ports and line.When the arterial lime was switched, the waveform was accurate.When a new catheter was used better readings were observed.Patient procedural time was extended as the numbers for diagnosis was altered.There was no patient injury.They used a new catheter that worked successfully.The transducer was discarded at the hospital.Patient demographics were unable to be obtained.
|
One 131f7 with a monoject limited volume syringe was returned for examination.The reported event of pressure issue was not confirmed during the evaluation.The pressure monitoring kit was not returned.All through lumens were patent without any leakage or occlusion.The balloon inflated clear and remained inflated for 5 minutes without leakage.The thermistor was found to read 37.0 c when submerged into a 36.9 c water bath.No visible damage or abnormality was observed from the catheter body or syringe.The reported event could not be confirmed or replicated during the analysis, as the device responded appropriately during functional testing.It is not known if some procedural factors may have contributed to the event.There was no evidence of a manufacturing nonconformance.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications are well described in the literature.With any hemodynamic monitoring, pressure readings can change quickly and dramatically.Pressure tubing that is used with swan ganz catheters can also be a contributing factor to inaccurate values.In regard to the pressure tubing used with swan ganz catheters, it should be noted that poor dynamic response can be caused by air bubbles, clotting, excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.Pressure readings should correlate with the patient¿s clinical manifestations.It is unknown whether user or procedural factors contributed to the stated event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|