Model Number 580220101 |
Device Problem
Mechanical Jam (2983)
|
Patient Problems
Injury (2348); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/13/2020 |
Event Type
Injury
|
Event Description
|
It was reported that during a 3-level case (l2-l5), two (2) tritanium x tl draw rods were jammed/ damaged in the inserter intra-operatively.The first level (l4/5) implant was placed as planned, achieving the desired expansion and final placement.Upon implant insertion of the second level (l3/4), the inserter¿s draw rod would not detach from the implant.The surgeon used a slap hammer to disengage the implant but was unable to.The implant was then removed, un-collapsed (expanded to approximately 1mm) with the inserter.A new inserter was then assembled, and a new implant was attached.Based on the anteroposterior x-rays, the implant was placed slightly past midline.The surgeon attempted to re-attach the inserter to adjust the cage placement closer to midline but was unable to re-attach.The implant was left as is with satisfactory positioning.The inserter was tested on the back table where it was discovered that the draw rod was damaged and not able to be disassembled from the inserter.Subsequently, the third level (l2/3) was completed with an alternate static titanium implant system.There were no adverse consequences to the patient and surgery was successfully completed with a reported surgical delay of 1.5 hours.This record represents the first of the two draw rods.
|
|
Manufacturer Narrative
|
Device lot number has been received and added to section d.
|
|
Event Description
|
It was reported that during a 3-level case (l2-l5), two (2) tritanium x tl draw rods were jammed/ damaged in the inserter intra-operatively.The first level (l4/5) implant was placed as planned, achieving the desired expansion and final placement.Upon implant insertion of the second level (l3/4), the inserter¿s draw rod would not detach from the implant.The surgeon used a slap hammer to disengage the implant but was unable to.The implant was then removed, un-collapsed (expanded to approximately 1mm) with the inserter.A new inserter was then assembled, and a new implant was attached.Based on the anteroposterior x-rays, the implant was placed slightly past midline.The surgeon attempted to re-attach the inserter to adjust the cage placement closer to midline but was unable to re-attach.The implant was left as is with satisfactory positioning.The inserter was tested on the back table where it was discovered that the draw rod was damaged and not able to be disassembled from the inserter.Subsequently, the third level (l2/3) was completed with an alternate static titanium implant system.There were no adverse consequences to the patient and surgery was successfully completed with a reported surgical delay of 1.5 hours.This record represents the first of the two draw rods.
|
|
Event Description
|
It was reported that during a 3-level case (l2-l5), two (2) tritanium x tl draw rods were jammed/ damaged in the inserter intra-operatively.The first level (l4/5) implant was placed as planned, achieving the desired expansion and final placement.Upon implant insertion of the second level (l3/4), the inserter¿s draw rod would not detach from the implant.The surgeon used a slap hammer to disengage the implant but was unable to.The implant was then removed, un-collapsed (expanded to approximately 1mm) with the inserter.A new inserter was then assembled, and a new implant was attached.Based on the anteroposterior x-rays, the implant was placed slightly past midline.The surgeon attempted to re-attach the inserter to adjust the cage placement closer to midline but was unable to re-attach.The implant was left as is with satisfactory positioning.The inserter was tested on the back table where it was discovered that the draw rod was damaged and not able to be disassembled from the inserter.Subsequently, the third level (l2/3) was completed with an alternate static titanium implant system.There were no adverse consequences to the patient and surgery was successfully completed with a reported surgical delay of 1.5 hours.This record represents the first of the two draw rods.
|
|
Manufacturer Narrative
|
The visual inspection concluded that the instruments were received with the inner shaft jammed in the inserter.Implant was attached to the inserter as it was unable to be disassembled during/after the surgery due to inner shaft jamming.During the functional inspection it was not possible to disengage the implant/inner shaft from the inserter.In order to disassemble, the tip of the inner shaft was cut off in r&d lab which allowed the inner shaft to be disassembled.Upon removal, the u joint ears were deformed and measured 0.2" on one side and 0.0134 one the opposite side, nominally 0.180" (measured with comparator).The inner shaft is no longer functional.Device history records were reviewed for this lot , no relevant manufacturing discrepancies were found.Complaint history records were reviewed for this lot, no similar complaints were identified.A corrective and preventative action was opened to identify the root cause of the deformation of the draw rod and subsequent jamming during use to prevent the recurrence of the issue.Per what was reported, the inserter was lubricated prior to use and the surgical technique was followed throughout the procedure.The inserter was not twisted or cantilevered, the inserter was unlocked during disassembly from the implant and during removal from disc space, the inserter was locked during impaction, while attaching the implant laser mark on the inserter denoted ¿lateral¿ was aligned with the laser mark on the graft block denoted "lateral", there was no cross threading between the implant/tube set/inserter during assembly.Furthermore it was reported that the disc space was tight and inserter was excessively impacted to place implant into final position (impaction force was rated as 4.5-5 on a scale 1-5 with 5 being excessive).9 mm cage was used, the implant was not undersized based off the trial height as 8 mm cage was not available.It is possible that excessive force applied to insert the cage into tight disc space contributed to binding between the cage and the inserter and difficulty with disassembling the inner shaft from the inserter.Per r&d analysis there is no evidence that implant sizing would impact these failures.Every size of implant is taller than the instrumentation including the junction block and tube set so even if the implant was a tight fit, the inserter features shouldn't be seeing the same frictional forces and bony anatomy that the implant is.A tight-fitting implant would require more impaction, however the impaction is most likely not a root cause of the failures.
|
|
Event Description
|
It was reported that during a 3-level case (l2-l5), two (2) tritanium x tl draw rods were jammed/ damaged in the inserter intra-operatively.The first level (l4/5) implant was placed as planned, achieving the desired expansion and final placement.Upon implant insertion of the second level (l3/4), the inserter¿s draw rod would not detach from the implant.The surgeon used a slap hammer to disengage the implant but was unable to.The implant was then removed, un-collapsed (expanded to approximately 1mm) with the inserter.A new inserter was then assembled, and a new implant was attached.Based on the anteroposterior x-rays, the implant was placed slightly past midline.The surgeon attempted to re-attach the inserter to adjust the cage placement closer to midline but was unable to re-attach.The implant was left as is with satisfactory positioning.The inserter was tested on the back table where it was discovered that the draw rod was damaged and not able to be disassembled from the inserter.Subsequently, the third level (l2/3) was completed with an alternate static titanium implant system.There were no adverse consequences to the patient and surgery was successfully completed with a reported surgical delay of 1.5 hours.This record represents the first of the two draw rods.
|
|
Manufacturer Narrative
|
Stryker initiated a voluntary recall on 17 march 2021 and an urgent medical device recall letter was sent to the affected customers notifying them of the product issue and associated risks and providing instructions pertaining to the removal and return of the product to stryker.
|
|
Search Alerts/Recalls
|
|