DEPUY ORTHOPAEDICS INC US PFC*SIGMA C/S(PS)BOX TRIAL SZ5; KNEE INSTRUMENT : FEMORAL TRIALS Back to Search Results
Model Number 96-1045
Device Problems Use of Device Problem (1670); Component Missing (2306); Device-Device Incompatibility (2919)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The screw, which allows to maintain the test block, does not have a stopper as usual.Once the screw goes out, it was impossible to replace it.This issue occurred on several devices.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported damage.The root cause is attributed to improper user technique, and no corrective action is indicated at this time.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
KNEE INSTRUMENT : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10531839
MDR Text Key206930800
Report Number1818910-2020-20020
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10603295234739
UDI-Public10603295234739
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 08/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number96-1045
Device Catalogue Number961045
Device Lot NumberSO2041402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Date Manufacturer Received11/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-