• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Defective Component (2292)
Patient Problem No Patient Involvement (2645)
Event Date 08/20/2020
Event Type  malfunction  
Manufacturer Narrative
An olympus field service engineer (fse) replaced the grey connector on site.The device was repaired according to regulations.No other issues were reported.If additional information is obtained a supplemental report will be filed.
 
Event Description
The user facility reported that a grey connector needed to be replaced.The issue was found during reprocessing.No patient involvement or injuries were reported.No additional information has been obtained.
 
Manufacturer Narrative
The investigation has been completed.A device history record (dhr) review was performed and found no non-conformances that would cause the reported malfunction.All records indicate that the device was manufactured in accordance with all applicable procedures.The instructions for use (ifu) states: ¿check the following for each connector.The connector should be fixed firmly.The o-rings should be free of abnormalities such as cracks, tears, or dents.¿ the root cause could not be conclusively determined.Probable causes that could have led to the reported event include that stress/impact was added to the connector toward loose direction by the user handling, and the accumulated stress caused the breakage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10531951
MDR Text Key222096228
Report Number8010047-2020-06493
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/20/2020
Initial Date FDA Received09/14/2020
Supplement Dates Manufacturer Received11/18/2020
Supplement Dates FDA Received12/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-