Model Number 22-4035 |
Device Problems
Break (1069); Material Separation (1562)
|
Patient Problems
No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/21/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that, during surgery, the upper jaw of the "firstpass suture passer" detached from the handle and shaft whilst inside the patient.The detached piece was successfully retrieved from the patient.The procedure was successfully completed without significant delay using a back-up device.No further complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined.A review of the device history records for the device showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A review of risk management files found that the reported failure was documented appropriately.A complaint history review concluded this was a repeat issue.A review of the instructions for use found that mishandling the device can result in failure.Without the reported product a visual and functional evaluation cannot be performed and customer¿s complaint cannot be confirmed.Factors that could have contributed to the reported event include: (1) excessive force.No containment or corrective actions are recommended at this time.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
|
Manufacturer Narrative
|
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows a detached suture capture.No manufacturing abnormalities.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) excessive force (2) tissue thickness (3) damage or debris on the device tip between passes.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|