Model Number 10848600
Device Problems Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/28/2020
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis one system.During a cardiology procedure, the user reported that the monitor display in the exam room started to blink (image appearing and disappearing).The user restarted the system ,but couldn't fix the problem.The patient was transferred to another system in the same hospital, and finished the procedure successfully.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The investigation was performed considering complaint description, cs reports, system history and system log files.The investigation showed that the cause of the error was due to a graphic card defect.According to information from field, during s cardiology procedure the monitor display in the exam room started to blink (image appearing and disappearing).The customer attempted to restart the system but couldn't fix the problem.The customer service engineer (cse) on site checked and replaced the graphic card of the smart pc.After the hardware exchange, the system recovered and works as specified.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ONE
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS SHENZHEN MAGNETIC RESONANCE LTD.
gaoxin c ave.
2nd, hi-tech industrial park
shenzhen, china 51805 7
CH  518057
MDR Report Key10534046
MDR Text Key206966059
Report Number3004754211-2020-44310
Device Sequence Number1
Product Code OWB
UDI-Device Identifier04056869009964
UDI-Public04056869009964
Combination Product (y/n)N
PMA/PMN Number
K133580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Repair
Type of Report Initial,Followup
Report Date 09/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10848600
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-