• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC THREE-WAY PLASTIC STOPCOCK; KGZ ACCESSORIES, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC THREE-WAY PLASTIC STOPCOCK; KGZ ACCESSORIES, CATHETER Back to Search Results
Model Number G00219
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/31/2020
Event Type  malfunction  
Manufacturer Narrative
Pma/510k #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
As reported, prior to an unknown procedure a three-way plastic stopcock was being prepared with surgical glue when it was found to leak.Another three-way plastic stopcock was then opened and being prepared in similar fashion when it, too, was found to leak.It is unknown how the procedure was completed.No portion of the device was left inside the patient.This did not result in additional procedures or prolonged hospitalization.The patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Description of event: as reported, prior to an unknown procedure a three-way plastic stopcock was being prepared with surgical glue when it was found to leak.Another three-way plastic stopcock was then opened and being prepared in similar fashion when it, too, was found to leak.It is unknown how the procedure was completed.Investigation / evaluation: a visual inspection and functional testing of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record, documentation, drawings, the instructions for use, manufacturing instructions, quality control data, and specifications.Two stopcocks received prior to use.Inspection found the devices are cracked.The two devices returned are for 308549 (qty 1) and 308550(qty 1).A review of the device history record found no non-conformances related to the reported failure mode.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.A trackwise search revealed one other complaints associated to the lot number: 8016146.It was reported by the same customer for the same failure mode, reported under (b)(4).A capa was opened due to consistent complaint history of this product surrounding cracking and subsequent leaking of stopcocks.The root cause of cracking / leaking stopcocks was determined to be stopcock stems molded of nylon 6 have the potential to absorb moisture resulting in an increase in diameter raising the potential for cracking of the stopcock body.Based on this result of the clinical effects analysis, the decision was made to no longer pursue corrective actions within capa.Cook has concluded that the conclusion from previously completed root cause investigation aligns with the conclusion of this complaint.The stopcock stems molded of nylon 6 have the potential to absorb moisture resulting in an increase in diameter raising the potential for cracking of the stopcock body contributed to this incident.Per the quality engineering risk assessment, no further action is warranted.Cook will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No new patient or event information to report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE-WAY PLASTIC STOPCOCK
Type of Device
KGZ ACCESSORIES, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key10534245
MDR Text Key206962852
Report Number1820334-2020-01686
Device Sequence Number1
Product Code KGZ
UDI-Device Identifier00827002002197
UDI-Public(01)00827002002197(17)220623(10)8016146
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/23/2022
Device Model NumberG00219
Device Catalogue NumberPTWS-2FLL-MLL-R
Device Lot Number8016146
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Date Manufacturer Received12/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-