Model Number 10603
Device Problem Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 03/18/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr.: synergy ii us mr 2.25 x 20mm stent delivery system was returned for analysis.A visual examination of the stent found signs of stent damage.A stent strut in the most proximal strut row was lifted and pulled in a distal direction.The undamaged stent outer diameter was measured and the result is within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no issues.A visual and tactile examination of the hypotube found multiple kinks along the length of the hypotube shaft.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 20aug2020.It was reported that shaft kink occurred.A 2.25 x 20 synergy drug-eluting stent was selected for use.However, it was noted that the stent was kinked and could not be used in the procedure.The procedure was completed with a different device.No patient complications nor injuries reported.However, returned device analysis revealed a stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Type of Device
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10536517
MDR Text Key207132022
Report Number2134265-2020-11648
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08714729840145
UDI-Public08714729840145
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/29/2021
Device Model Number10603
Device Catalogue Number10603
Device Lot Number0024570876
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-