Model Number 2549-00-070
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/24/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation summary: examination of the returned device confirmed the reported event.The coating on the brackets was delaminating.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Event Description
It was reported that during post cleaning inspection the pan was found to be damaged.The plastic holder for the shims was broken off.No surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REV CONV&OFFST FEM PREP
Type of Device
INSTRUMENT CASES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key10537755
MDR Text Key207052384
Report Number1818910-2020-20114
Device Sequence Number1
Product Code FSM
UDI-Device Identifier10603295476016
UDI-Public10603295476016
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial
Report Date 08/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2549-00-070
Device Catalogue Number254900070
Device Lot NumberPV0156626
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2020
Date Manufacturer Received09/02/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-