Model Number 71441265
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/17/2020
Event Type  Injury  
Event Description
It was reported that during surgery, while outside the patient, device was noticed to have the plastic gasket broken.Procedure was concluded using a double sided spacer block.There was no injury or delay.
 
Manufacturer Narrative
Ave broken.Per complaint details, the procedure was concluded using a double sided spacer block.There was no patient injury or delay reported.Therefore, since no patient harm is alleged, no further clinical assessment is warranted.A visual inspection confirmed the plastic rings are broken on the gii spc pri spacer block.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII SPC PRI SPACER BLOCK
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10539585
MDR Text Key207102584
Report Number1020279-2020-04607
Device Sequence Number1
Product Code HRY
UDI-Device Identifier00885556034354
UDI-Public00885556034354
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71441265
Device Catalogue Number71441265
Device Lot Number07KM10263
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2020
Initial Date Manufacturer Received 08/18/2020
Initial Date FDA Received09/16/2020
Supplement Dates Manufacturer Received02/04/2021
Supplement Dates FDA Received02/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-