• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 5.5X36MM PERIPHERAL SCREW, LOCKING; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 5.5X36MM PERIPHERAL SCREW, LOCKING; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number 5.5X36MM PERIPHERAL SCREW, LOCKING
Device Problem Mechanics Altered (2984)
Patient Problem Impaired Healing (2378)
Event Date 08/25/2020
Event Type  Injury  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event Description
It was reported that (b)(6) 2019 a patient had a reverse shoulder procedure.On (b)(6) 2020 the patient underwent a revision surgery.The patient had an autograft to the glenoid on her initial surgery and it did not appear to have healed.The sales representative who was present during the revision reports that the surgeon had mentioned in surgery that he believed that the locking mechanism on the baseplate may not have been functioning appropriately (the screw was not locked in).During the original procedure the following arthrex devices had been implanted: ar-9564-2436-lat, glenosphere, lot 18.00689.Ar-9503s-03, humeral insert, lot 18.00870.Ar-9502f-36cpc, univers revers suture cup, lot 170141708.Ar-9561-25p, modular central post, lot 5368.Ar-9560-24-2, modular baseplate, lot 5408.Ar-9562-36nl, peripheral screw non-locking, lot 2018001207.Ar-9563-36, peripheral screw locking, lot 2018001490.Ar-9563-40, peripheral screw locking, lot 2018001491.Ar-9501-12p, univers revers humeral stem size 12, lot 170125705.During the revision surgery all off the above listed original implants were explanted, except for the ar-9501-12p, univers revers humeral stem size 12, which was left in the patient.Explanted devices will be returned for evaluation.The surgeon completed the revision by doing a hemi arthroplasty and implanting the following arthrex products: ar-9502-36arca, revers ca adapter assembly, lot 18.01305.Ar-9544-17rca, revers ca humeral head, lot 10071302.Ar-9502f-36cpc, univers revers suture cup, lot 19.00846.The original and revision procedures were performed by the same surgeon at the same facility.
 
Manufacturer Narrative
Complaint not confirmed, despite damage to the locking thread the device was found to lock in the baseplate.The device was found to meet dimensional specifications.The cause of the event is undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5X36MM PERIPHERAL SCREW, LOCKING
Type of Device
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key10540490
MDR Text Key207297967
Report Number1220246-2020-02165
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888867296909
UDI-Public00888867296909
Combination Product (y/n)N
PMA/PMN Number
K173900
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Model Number5.5X36MM PERIPHERAL SCREW, LOCKING
Device Catalogue NumberAR-9563-36
Device Lot Number2018001490
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-