DEPUY IRELAND - 9616671 ATTUNE FB CR ART SURF SZ9 CONV; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Model Number 2545-01-969
Device Problem Break (1069)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Date 09/01/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Has product been potentially contaminated or is contaminated with bodily fluids? yes.2545-00-928 has a crack that was discovered during cleaning/steam sterilization.2545-01-969 has broken into 2 pieces that happened per-op.2544-00-502 a piece of plastic fell off during use per-operative.Where / when did the event occur? intra-op, post - op what action was taken/required to manage the problem during the procedure? with same like product.
 
Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported damage.The noted damage is consistent with material overload through the use of excessive force and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB CR ART SURF SZ9 CONV
Type of Device
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10541469
MDR Text Key207169503
Report Number1818910-2020-20221
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295423416
UDI-Public10603295423416
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2545-01-969
Device Catalogue Number254501969
Device Lot NumberBFA0WFH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2020
Initial Date Manufacturer Received 09/02/2020
Initial Date FDA Received09/16/2020
Supplement Dates Manufacturer Received09/23/2020
10/01/2020
Supplement Dates FDA Received09/30/2020
10/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-