Model Number 7210812S
Device Problem Failure to Power Up (1476)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
It was reported that the vulcan generator had a powe drops.No case involved.Therefore, there was no patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10, h2, h3, h6: the device, intended for use, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of power output failure could not be reproduced.Unit passed all functional testing with no loss of power.All functions perform as expected.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL VULCAN GENERATOR CE MARK
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10543382
MDR Text Key207232929
Report Number1643264-2020-01177
Device Sequence Number1
Product Code GEI
UDI-Device Identifier03596010542182
UDI-Public03596010542182
Combination Product (y/n)N
PMA/PMN Number
K991140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7210812S
Device Catalogue Number7210812S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2020
Initial Date Manufacturer Received 08/19/2020
Initial Date FDA Received09/16/2020
Supplement Dates Manufacturer Received10/12/2020
Supplement Dates FDA Received10/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-