|
Model Number SPD2-US-060-320 |
Device Problems
Peeled/Delaminated (1454); Deformation Due to Compressive Stress (2889)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluation visual inspection: an inspection of the spider fx showed no damages or anomalies to the filter assembly or green distal segment of the capture wire.An inspection of the black proximal segment of the capture wire showed the areas of the wire where ptfe was not present.Areas of the exposed wire were noted at the following areas measured from the proximal tip of the capture wire: 11-13cm, 21-33cm, and 35-38cm.Under microscope the areas of compromised ptfe are longitudinal and is not present around the circumference of the wire.An area of the exposed wire shows the pft curved outward and then back proximally at a distal edge of the observed damage and string-like debris was observed.It is possible the debris may be fibrous from an external source used to wipe the capture wire.The pffe shows characteristics consistent with exposure to friction, possibly scraped.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Physician was attempting to use a spider fx embolic protection during procedure.It was reported that, during the procedure, as the surface coating of the spider wire was peeled off, there was a defect of sedimentation of coating foreign matter in the catheter and wire kinking.It is unknown if the spider wire was exposed when the hydrophilic coating peel was noted.The device was removed safely and the procedure was completed with a new medtronic device.There was no patient injury reported.When the device was returned to the manufacturing facility, it was noted confirmed the coating was peeled.
|
|
Search Alerts/Recalls
|
|
|