Model Number 10674 |
Device Problem
Break (1069)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
|
Event Date 08/22/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that shaft break occurred.A 24 x 2.50 promus elite drug-eluting stent was selected for use.However, during preparation, the catheter of the stent delivery system was found to be broken.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.
|
|
Event Description
|
It was reported that shaft break occurred.A 24 x 2.50 promus elite drug-eluting stent was selected for use.However, during preparation, the catheter of the stent delivery system was found to be broken.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.It was further reported that the device was broken during removal from the carrier coil tube.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: a promus elite ous mr 24 x 2.50mm stent delivery system was returned for analysis without the carrier hoop.An examination of the crimped stent found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon was reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found a break 2mm distal to the distal end of the strain relief.Examination of the break site via scope found that the hypotube appeared to have been bent at the break site, as evident by the curved oval shaped appearance of the hypotube on both sides of the break.An examination of the shaft polymer extrusion found no issues.An examination of the tip found no issues.No other issues were identified during the product analysis.
|
|
Search Alerts/Recalls
|