Model Number 2217-50-041 |
Device Problem
Break (1069)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Code Available (3191)
|
Event Date 09/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
During corail/pinnacle surgery the cup impactor broke.The hcp was hitting with the hammer the cup impaction handle when the end of the handle fell to the floor.The surgery wasn¿t affected by this incident because we had a spare on site.
|
|
Manufacturer Narrative
|
Product complaint # = > (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).Investigation summary : examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
|
|
Search Alerts/Recalls
|