Model Number 1458Q/86 |
Device Problems
Difficult to Insert (1316); Separation Failure (2547); Device Dislodged or Dislocated (2923)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product is registered as a combination product.The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
It was reported that when the patient presented for implant procedure, the left ventricular lead dislodged after implanted at suitable position.During the second implant attempt, the guidewire was difficult to insert into the lead and could not be separated back after insertion.The lead was not used and replaced.The patient was stable.
|
|
Manufacturer Narrative
|
The reported event of guidewire difficult to insert/remove was confirmed.A complete lead was returned for analysis with a broken guidewire inserted in the inner coil lumen.The cause of the reported event of guidewire insertion and removal difficulty was due to guidewire being bent/kinked, inner coil offset and ptfe coating of the guidewire stripped.All damages found were sustained during the surgical procedure.The lead was otherwise normal.
|
|
Search Alerts/Recalls
|