Model Number 7008605
Device Problems No Display/Image (1183); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2020
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the axiom artis dta system.During an interventional procedure, the user reported that no images were displayed after x-ray acquisition.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware issue.The investigation showed that the error was caused by an image system issue.The file system of the image system computer (rt-pc) went corrupt leading to the error.If such an error occurs, the designed mitigation is to switch to "bypass fluoro mode".When the imaging system is not available, as in the given case, the system enters this mode where only continuous fluoroscopy with compromised image quality is available and the acquired scenes cannot be saved and reviewed.Normal system functionality can be restored only after local service intervention.The service engineer went onsite and restored the data base which corrected the file system and eliminated the error.Afterwards, the system was returned to operation and the issue has not been reported again.A general error which would require corrective action could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOM ARTIS DTA
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH- AT
siemensstrasse 1
forchheim, germany 91301
GM  91301
MDR Report Key10553562
MDR Text Key207560759
Report Number3004977335-2020-45533
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
PMA/PMN Number
K052202
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Repair
Type of Report Initial,Followup
Report Date 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7008605
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-