Model Number MC2-4590R
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Failure of Implant (1924)
Event Date 08/24/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to manufacturer for investigation.Root cause is unable to be determined.Device was returned to bfarm.
 
Event Description
Information was received that a revision procedure was performed on (b)(6) 2020.As per the reporter x-ray images revealed the end cap was disengaged from the actuator.As per the surgeon there is a possibility for infection.
 
Manufacturer Narrative
Even though no product has been returned, fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
 
Event Description
Information was received that a revision procedure was performed on august 24, 2020.As per the reporter x-ray images revealed the end cap was disengaged from the actuator.As per the surgeon there is a possibility for infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key10557804
MDR Text Key207696890
Report Number3006179046-2020-00394
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00887517944313
UDI-Public887517944313
Combination Product (y/n)N
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Remedial Action Recall
Type of Report Initial,Followup
Report Date 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMC2-4590R
Device Lot Number8062118
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1898-2020
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
-
-