It was reported that during tendoachilles repair procedure, halfway into insertion of the anchor into the bone, it broke into 2 pieces, it was successfully removed from the patient using forceps.Delay was grater than 30 minutes and it is unknown how the procedure was completed.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H10 h3, h6: part of the reported device was received for evaluation.There was no relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.A visual inspection of the returned device found that it is not in its original packaging.The anchor and sutures were not returned with the device.The distal end of the inserter is bent, and there is debris on the device.It was reported that during a tendoachilles repair procedure, the anchor broke into two pieces halfway into insertion.Based on the condition of the product material found during visual inspection, additional material testing is not required.Per the complaint details, all pieces were removed from the patient using forceps.There was a surgical delay of greater 30 minutes but it is unknown how the procedure was completed.No further complications were reported.Per communication via e-mail, the patient¿s current status is normal.Therefore, no further clinical/medical assessment is warranted at this time.The complaint was not confirmed, and the root cause could not be determined.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|