• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Failure to Charge (1085); Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 21sep2020.
 
Event Description
It was reported that the ventilator was running on the external battery even though it was plugged into alternating current (ac) power.Reportedly, the unit would not charge the battery.There was no patient involvement.The customer troubleshot with technical support and reported that there was no 24 volt from the power supply.
 
Manufacturer Narrative
G4: 04nov2020.B4: 05nov2020.The customer had been advised when troubleshooting with technical support to confirm 24 volt supply from power supply by checking the power supply output.Upon a follow up with technical support the customer reported part had been ordered and it would be installed.No additional information was provided.Multiple attempts were made to obtain information on the repair/service of the device that have been unsuccessful.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10561106
MDR Text Key208523491
Report Number2031642-2020-03370
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/01/2020
Initial Date FDA Received09/21/2020
Supplement Dates Manufacturer Received09/01/2020
Supplement Dates FDA Received11/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-