Submission of this report does not constitute an admission that the manufacturer's product caused, or contributed to the event.Device evaluation by manufacturer: a distributor engineer visited the customer's site to address the reported issue.Fse performed maintenance to the waste pump, and confirmed it was functioning properly.Fse also shut down the analyzer, cleaned, and dried the area of the analyzer.The customer successfully performed quality control (qc) without any errors.The aia-360 analyzer returned to operation.No further action required by field service.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-360 operator's manual under section 7-1 list of error messages states the following: leak sensor s702 detected is generated when the leakage sensor s702 activated.The operator is instructed to contact the service department.The most probable cause of the reported event was due to waste pump needed maintenance.
|
A customer reported getting error message "3022 leak sensor s702 detected" on the aia-360 analyzer.A distributor engineer instructed the customer to take the upper and back cover of the analyzer off, and inspect the s702 sensor, customer confirmed leakage in the sensor area.A distributor engineer was dispatched to address the reported event, which resulted in delayed reporting of patient results for alpha-fetoprotein (afp), prolactin (prl), and progesterone (prog ii).There was no indication of any patient intervention, or adverse health consequences due to the delay in reporting of patient results.
|