MAUDE Adverse Event Report: TOSOH CORPORATION AIA-360

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION AIA-360 Back to Search Results
Model Number AIA-360
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation by manufacturer: a field service engineering (fse) was at the customer's site to resolve reported event.Fse confirmed reported error by reviewing analyzer print out.Fse was able to reproduce the reported event by running a substrate replacement.Fse resolved the complaint by replacing the 3 way solenoid valve; fse also replaced the syringe tip.Fse performed quality control run to validate the analyzer, run completed without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review, and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-360 training manual, tab 8: troubleshooting states the following: [2017] substrate purge failure is generated when no substrate was dispensed at daily maintenance.The operator is instructed to check substrate level and replace as necessary, repeat daily maintenance the most probable cause of the reported event is due to faulty 3-way solenoid valve.
 
Event Description
A customer reported getting error message "2017 substrate purge failure" during aia-360 analyzer shutdown.Technical support specialist (tss) instructed the customer to put fresh alcohol in the bottle and try to do a replace substrate from the maintenance, operation failed with error 2017 reoccurring.Customer attempted shutdown again, error persisted.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for intact parathyroid hormone (ipth).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
AIA-360
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA   1058623
Manufacturer Contact
bernadette oconnell
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586-23
JA   1058623
MDR Report Key10568168
MDR Text Key208013117
Report Number8031673-2020-00277
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360
Device Catalogue Number019945
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-