On (b)(6) 2020, a 35mm amplatzer pfo occluder was selected for implant.During the procedure, while the device was deployed inside the patient's body, the ra disc formed into a bowl shape.The device was retracted and removed from the patient.A new 30mm amplatzer pfo occluder was successfully implanted.The patient is reported to be in stable condition.
|
Additional information sections: d10, h2, h3, h6, h10.The reported event of a round, bowl shape deployment was confirmed.Following the simulated deployment, the bulbous shape returned to normal and met functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.A capa was initiated for further investigation and did not identify a product quality issue.However, corrective actions to further enhance performance are being pursued per internal operating procedures.
|