Visual analysis was performed on the returned device.The reported separation was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation determined that the reported separation was related to circumstances of the procedure.It is likely that the bends and kinks noted in the shaft occurred during advancement.Additionally, after successful deployment of the stent, it is likely that the kinked locations of the hypotube separated as the sess was being removed and the kinked hypotube material was straightened.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
|