• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER; TEMPERATURE SENSING CATHETER (SILICONE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER; TEMPERATURE SENSING CATHETER (SILICONE) Back to Search Results
Model Number 119108
Device Problems Material Fragmentation (1261); Component Missing (2306)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the tip of the catheter had been missing when the package was opened.
 
Event Description
It was reported that the tip of the catheter had been missing when the package was opened.
 
Manufacturer Narrative
The reported event was confirmed however the cause was unknown.Visual inspection noted one temperature sensing catheter was received.Visual evaluation noted that the catheter did not have a complete tip on return.The catheter appears to have an angled cut, the removed the tip.The eyelets were not present on the device.This fails to meet specifications as voids and incomplete tips are not acceptable.Although the reported event was confirmed, a root cause could not be determined.A potential root cause for this failure could be tube contamination.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Sterile unless package is opened or damaged.Warning: do not use ointments or lubricants having a petrolatum base.They will damage silicone and may cause balloon to burst.Caution: do not aspirate urine through the drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip syringe.Do not use needle.Recommended inflation capacities.8 fr.And 10 fr.(3cc balloon): use 3.5cc sterile water.12 fr.(5cc balloon): use 5.5cc sterile water.14 fr.And larger (5cc balloon): use 10cc sterile water.Do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Caution: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone balloon foley catheters.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Note: compatible only with the bard criticore monitor, bard urotrack 224 monitor, and other 400-series temperature monitors.Interchangeability + 0.2oc at 37oc.Warning: this product should never be connected to the temperature monitor or connected to a cable during an mri procedure.Failure to follow this guideline may result in serious injury to the patient.It is important to closely follow these specific conditions that have been determined to permit the examination to be conducted safely.Any deviation may result in a serious injury to the patient.Caution: as with all temperature probes, in the presence of rf energy sources, local heating, temperature errors, and probe damage may occur.Do not stretch catheter.This will cause repositioning of probe.Do not use stylet.This will cause stretching of catheter." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER
Type of Device
TEMPERATURE SENSING CATHETER (SILICONE)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10575261
MDR Text Key208750999
Report Number1018233-2020-06172
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741045998
UDI-Public(01)00801741045998
Combination Product (y/n)N
PMA/PMN Number
K070582
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number119108
Device Catalogue Number119108
Device Lot NumberNGEP0705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2020
Date Manufacturer Received12/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-