• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION Back to Search Results
Model Number RA002-5555SL
Device Problem Defective Device (2588)
Patient Problem Failure of Implant (1924)
Event Date 02/15/2016
Event Type  Injury  
Manufacturer Narrative
The rod was received for visual and functional testing and the reported event was confirmed as the device was jammed.The root cause of the reported event may be related to bending force applied during the distraction sessions and/or patient's daily activities.The distraction rod rubbing against the housing tube from the patient's activity would cause debris buildup, which would cause the device to jam.Review of the device history record for the returned rod and confirmed that it met all of the required quality inspections.
 
Event Description
Information was received that a revision procedure was performed on (b)(6) 2016.As per the reporter, the rods were removed for converting to final fusion and appeared to be jammed.The patient successfully completed treatment and lengthening sessions without issues; however, during the last distraction session, the rod did not appear to lengthen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED
101 enterprise
suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED
101 enterprise
suite 100
aliso viejo CA 92656
Manufacturer Contact
tiara rae
101 enterprise
suite 100
aliso viejo, CA 92656
6197314188
MDR Report Key10575409
MDR Text Key208192805
Report Number3006179046-2020-00397
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00856719002190
UDI-Public856719002190
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRA002-5555SL
Device Lot NumberA130116-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age14 YR
Patient Weight38
-
-