Results: the jet7 was kinked and ovalized approximately 110.5 ¿ 112.5 cm from the hub.The total length of the jet7 was approximately 132.5 cm.Conclusions: evaluation of the returned jet7 was kinked and ovalized on the distal shaft.This damage was likely the reported stretching.If the device is forcefully manipulated against resistance, damage such as this may occur.During functional testing, the jet7 was advanced through a demonstration neuron max with resistance due to the ovalization.The root cause of initial resistance during the procedure could not be determined.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure in the left m1 segment of the middle cerebral artery (mca) using a penumbra system jet 7 reperfusion catheter (jet7), velocity delivery microcatheter (velocity), and non-penumbra sheath.During the procedure, the physician advanced the jet7 into the target vessel and turned on aspiration for approximately two minutes.The physician started to retract the jet7; however, resistance was encountered and the distal portion of the jet7 started to become stretched.Subsequently, the physician stopped retracting the jet7 and started retracting the sheath which in turn started to move the jet7.The physician then stopped retracting on the sheath and pulled the jet7 to remove it with the clot contained at the tip of the jet7.An angiogram was performed to reveal that the clot was removed from the artery.Post procedure, the physician mentioned a vasospasm occurred in the cervical internal carotid artery (ica) related to the sheath and, consequently, caused the jet7 to become stretched upon retraction.It was also reported that no medical intervention was performed to treat the vasospasm.There was no report of an adverse effect to the patient.
|