• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DISCOVERY ELBOW; DISC ULNA 2.5X53MM LT BOND CT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DISCOVERY ELBOW; DISC ULNA 2.5X53MM LT BOND CT Back to Search Results
Catalog Number 114802
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Unspecified Infection (1930)
Event Date 08/13/2020
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 52 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery - patient had irrigation & drainage for infection and condyle and bearing were exchanged.
 
Manufacturer Narrative
D10 - adding concomitant part 600-10-100.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC ULNA 2.5X53MM LT BOND CT
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key10580168
MDR Text Key208361101
Report Number1644408-2020-00822
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/28/2023
Device Catalogue Number114802
Device Lot Number577430
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/27/2020
Initial Date FDA Received09/24/2020
Supplement Dates Manufacturer Received08/27/2020
Supplement Dates FDA Received08/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
540-00-004 LOT 78341; 600-10-100 COBALT-G MV BONE CEMENT 40GM; 540-00-004 LOT 78341
Patient Outcome(s) Required Intervention;
Patient Age59 YR
Patient SexFemale
-
-