• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74122554
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Toxicity (2333); Limited Mobility Of The Implanted Joint (2671); Metal Related Pathology (4530)
Event Date 02/07/2019
Event Type  Injury  
Event Description
It was reported that a revision surgery was performed on the patients left hip on (b)(6) 2019, due severe pain, limited mobility, metallosis, elevated cobalt and chromium levels, and trunnionosis.Among the intra-operative findings and diagnoses were: elevated cobalt and chromium levels, trunnionosis and metallosis as a result of the premature failure of the devices.The patient's outcome is unknown.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the cup and hemi head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.Although the reported pain, elevated cobalt and chromium levels may be consistent with findings associated with trunnionosis and metal debris; however, without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the root cause of the pain, elevated cobalt and chromium levels, and trunnionosis cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 54MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key10582549
MDR Text Key208410185
Report Number3005975929-2020-00357
Device Sequence Number1
Product Code NXT
UDI-Device Identifier00885556071854
UDI-Public00885556071854
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date02/26/2013
Device Model Number74122554
Device Catalogue Number74122554
Device Lot Number08BW15618
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/28/2020
Initial Date FDA Received09/24/2020
Supplement Dates Manufacturer Received11/25/2020
Supplement Dates FDA Received12/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SLEEVE 74222200, LOT 08BW15618; STEM 71357107, LOT 07KM23182A; SLEEVE 74222200, LOT 08BW15618; STEM 71357107, LOT 07KM23182A
Patient Outcome(s) Hospitalization; Required Intervention;
-
-