• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC XR MP T1 PPS 15X115MM 115MM T1; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC XR MP T1 PPS 15X115MM 115MM T1; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 09/02/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(6).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported the circulated items in the warehouse identified debris in sterile packages.No patients were involved.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D4: (b)(4).Visual evaluation of the returned products identified that there is debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier and black debris consistent with porous material from the device.Review of the device history records identified no related deviations or anomalies.These products were likely conforming when they left zimmer biomet control.The root cause of the reported issue is attributed to transit damage causing the foam packaging to become abraded and shed.This reported event falls within the scope of a previous corrective action, the purpose of which is to assess all current sterile barrier systems used to package products at zimmer biomet bridgend.As part of this, the pouch is being improved to use a stronger material (nylon), and foam end caps are being added.Also, the orientation the devices are packed in the shipper box is moving from vertical to horizontal, and the thickness of the shipper box has been increased.Upon investigation, it has been determined that the debris in the sterile packaging meets the acceptable criteria and product is conforming to specifications.Event is no longer considered reportable, and initial report should be voided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
Upon further investigation, it has been determined that the debris is the sterile packaging meets the acceptable criteria specifications.This event is no longer considered reportable.Therefore, the initial report should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR MP T1 PPS 15X115MM 115MM T1
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10584506
MDR Text Key208588273
Report Number0001825034-2020-03615
Device Sequence Number1
Product Code OQG
Combination Product (y/n)N
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model NumberN/A
Device Catalogue Number51-145150
Device Lot Number2931560
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/02/2020
Initial Date FDA Received09/24/2020
Supplement Dates Manufacturer Received01/13/2021
Supplement Dates FDA Received01/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-