Model Number XS PST |
Device Problem
Circuit Failure (1089)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer¿s products have been requested for return.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.Occupation was lay user/patient.
|
|
Event Description
|
The initial reporter had an issue with the display of the coaguchek xs meter that could result in a misinterpretation of results.The initial reporter stated the segments in the results field of the display are partially fading.There was no allegation of any actual misinterpreted results.
|
|
Manufacturer Narrative
|
The customer's meter was received for investigation.The meter¿s circuit board was tested for damage or contamination and no contamination was found on the printed circuit board (pcb).The display shows no error during investigation (no missing or fainted segments).The circuit board shows no contamination that could temporarily lead to the complained error.Medwatch fields d9 and h3 were updated.
|
|