Model Number 27040XA
Device Problem Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 09/25/2020
Event Type  malfunction  
Manufacturer Narrative
The ceramic tip fractured/broke from the sheath.The sheath has a date code of june 2016 and the pictures show stress fracture from possible old age/use.
 
Event Description
Allegedly, during a turb procedure, the ceramic tip of the sheath broke off inside the patient's bladder.The doctor was able retrieve the broken piece with graspers.No harm to the patient; and procedure was completed with no other surgical intervention required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER CERAMIC INSULATION TUBE
Type of Device
RESECTOSCOPE SHEATH, INNER FIXED SHEATH
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen, gm
Manufacturer (Section G)
KARL STORZ SE & CO.KG
dr.-karl-storz -strasse 34
78532
tuttlingen, gm
Manufacturer Contact
susie chen
2151 e. grand avenue
el segundo, CA 90245
4242188201
MDR Report Key10586320
MDR Text Key209147269
Report Number9610617-2020-00105
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04048551076521
UDI-Public4048551076521
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K882270
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number27040XA
Device Catalogue Number27040XA
Device Lot NumberTT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/26/2020
Initial Date FDA Received09/25/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
Patient Weight84
-
-