Model Number CI-1400-01 |
Device Problems
Loose or Intermittent Connection (1371); Mechanical Problem (1384); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Therapeutic Response, Decreased (2271)
|
Event Date 12/01/2020 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing loss of lock despite the device testing results within normal limits.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The no lock condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device failed the residual gas analysis test.This device had moisture that exceeded the residual gas analysis test limit.Based on an assessment of the rga data, it is determined that this device was non-hermetic.This version of the 90k device is no longer distributed.Corrective actions were implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|