No product has been returned for evaluation as it remains in-situ, but provided radiograph confirmed alleged event.Even though root cause cannot be confirmed, review of the reported information identified a surgical technique may have contributed to alleged event.No revision procedure is planned at this time.Labeling review: "warning: care should be taken to insure that all components are ideally fixated prior to closure." "pre-operative warnings: care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." "note: check both reduction extensions after removal for the black lasermark at the distal end of the reduction extension to ensure the reduction extension has broken off at the correct spot note: be sure the lock screw is fully seated and the rod is fully locked down." device still in-situ.
|
On (b)(6) 2020, a patient underwent spinal procedure.On (b)(6) 2020, radiographs indicate the mass reduction screw did not break off correctly.As per reporter, it has not been decided if patient will undergo a revision procedure.
|