MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH CONFIDENCE SPINAL CEMENT SYSTEM CONFIDENCE PLUS KIT SPINAL CEMENT SYSTEM 11CC; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH CONFIDENCE SPINAL CEMENT SYSTEM CONFIDENCE PLUS KIT SPINAL CEMENT SYSTEM 11CC; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Model Number 283910000
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/01/2020
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a synthes employee.The instrument(s) was not returned, and instead, the investigation will be done based on the supplied image(s).The image(s) was reviewed, and the complaint condition for broken could be confirmed as the image provided shows the broken tube.As the instrument(s) was not returned an as received condition, dimensional inspection, material or drawing reviews are not applicable.A definitive assignable root cause could not be determined based on the provided information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from (b)(6) reports an event as follows: it was reported that on (b)(6) 2020, the plastic tubing broke during an unknown surgery.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.This report is for one (1) confidence spinal cement system confidence plus kit spinal cement system 11cc.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10: additional narrative: d4: lot number, d10; h3, h6: a product investigation was completed: visual inspection of the complaint device showed the tube connection end had broken apart.A dimensional inspection was not performed due to post-manufacturing damage.The relevant documents were reviewed; no design issues or discrepancies were identified.This complaint is confirmed as the tube connection end had broken apart.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIDENCE SPINAL CEMENT SYSTEM CONFIDENCE PLUS KIT SPINAL CEMENT SYSTEM 11CC
Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10587330
MDR Text Key208590753
Report Number1526439-2020-01802
Device Sequence Number1
Product Code NDN
UDI-Device Identifier10705034209623
UDI-Public10705034209623
Combination Product (y/n)N
PMA/PMN Number
K060300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283910000
Device Catalogue Number283910000
Device Lot Number281781
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2020
Date Manufacturer Received09/30/2020
Patient Sequence Number1
-
-