Model Number 8300 |
Device Problems
Break (1069); Crack (1135); No Device Output (1435); Naturally Worn (2988)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
This reported event and any subsequent repairs have been investigated through the normal service repair process.Failure data and parts used information, was reviewed for the sap file and track wise file, and found to be relevant to the service repair.A review of the device service history record was performed beginning from the date of manufacture to the present date and indicated that this device has been previously returned for service.Service returns were for issues similar to the reported complaint or similar to the service history.The database showed no quality notifications were opened for the source device.A review of the device history record in sap for sn (b)(4) was performed which confirmed that this device was not involved in a production failure and product was returned for the servicing which correlates to the customer reported issue.A review of the complaint history record in the trackwise was performed for the sn (b)(4) which confirmed no similar complaints with the same or related failure mode.Capa reference: (b)(4).The customer stated that there was no patient involvement.
|
|
Event Description
|
(b)(4).
|
|
Manufacturer Narrative
|
This reported event and any subsequent repairs have been investigated through the normal service repair process.Failure data and parts used information, was reviewed for the sap file and track wise file, and found to be relevant to the service repair.A review of the device service history record was performed beginning from the date of manufacture to the present date and indicated that this device has been previously returned for service.Service returns were for issues similar to the reported complaint or similar to the service history.The database showed no quality notifications were opened for the source device.A review of the device history record in sap for sn (b)(6) was performed which confirmed that this device was not involved in a production failure and product was returned for the servicing which correlates to the customer reported issue.A review of the complaint history record in the trackwise was performed for the sn (b)(6) which confirmed no similar complaints with the same or related failure mode.Capa reference : ca-2018-0171.The customer stated that there was no patient involvement.
|
|
Event Description
|
Possible oridion board issue.Oridion board- failed leak down test, case rear- damaged/cracked and iui- isolation rib damage.There was no patient involvement.01/19/201,8 08:16:49, (b)(6).Po for (b)(6) approved by (b)(6).Shipping & billing addresses confirmed.01/26/2018, 13:52:03 (b)(6).(b)(4).
|
|
Search Alerts/Recalls
|