• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAEMONETICS CORPORATION CELL SAVER ELITE PLUS AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 125ML,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAEMONETICS CORPORATION CELL SAVER ELITE PLUS AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 125ML, Back to Search Results
Model Number CSE-P-125-JA
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2020
Event Type  malfunction  
Manufacturer Narrative
On (b)(6) 2020, haemonetics manufacturing performed a visual inspection of the received bowl from the cell saver® elite set - 125ml, and confirmed blood in the inner core with a crack in the inner core base.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack, but did not cause, or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due the event of the past report.
 
Event Description
On (b)(6) 2020, haemonetics was notified of a returning error alarm, which was displayed on the 7th cycle during an aortic valve procedure in (b)(6), utilizing the cell saver® elite® autotransfusion system, and cell saver® elite set - 125ml.There was no reported impact to patients' health.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER ELITE PLUS AUTOTRANSFUSION SYSTEM
Type of Device
CELL SAVER ELITE SET - 125ML,
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer Contact
janice kiser
125 summer street
boston, MA 02110
MDR Report Key10591252
MDR Text Key208741407
Report Number1219343-2020-00083
Device Sequence Number1
Product Code CAC
UDI-Device Identifier10812747016537
UDI-Public(01)10812747016537(17)230320(10)0320056
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/20/2023
Device Model NumberCSE-P-125-JA
Device Lot Number0320056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-