HAEMONETICS CORPORATION CELL SAVER ELITE PLUS AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 125ML,
|
Back to Search Results |
|
Model Number CSE-P-125-JA |
Device Problems
Crack (1135); Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
On (b)(6) 2020, haemonetics manufacturing performed a visual inspection of the received bowl from the cell saver® elite set - 125ml, and confirmed blood in the inner core with a crack in the inner core base.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack, but did not cause, or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due the event of the past report.
|
|
Event Description
|
On (b)(6) 2020, haemonetics was notified of a returning error alarm, which was displayed on the 7th cycle during an aortic valve procedure in (b)(6), utilizing the cell saver® elite® autotransfusion system, and cell saver® elite set - 125ml.There was no reported impact to patients' health.
|
|