• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIVATOR II; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIVATOR II; SNARE, FLEXIBLE Back to Search Results
Model Number M00561221
Device Problems Retraction Problem (1536); Difficult to Open or Close (2921); Material Twisted/Bent (2981); Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/02/2020
Event Type  malfunction  
Manufacturer Narrative
The complainant was unable to provide the device lot number.Therefore, the manufacture and expiration dates are unknown.However, the complainant stated that the device was used prior to the expiration date.(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a captivator ii-10mm round stiff snare was used in the colon during a colonoscopy procedure performed on (b)(6) 2020.According to the complainant, during the procedure and inside the patient, when testing the snare prior to insertion, the snare did extend and retract but had a slight resistance and did not open and shut as smoothly as normal.After insertion into the scope and lining up beside the polyp, the snare would not sit or lie flat on the mucosal surface since the wire appeared to be curved.Additionally, the snare did not close well, then jumped through the tissue and cut without normal control.Reportedly, there were no issues noted with the handle.The snare was able to extend/retract after it was removed from the scope/patient but there was abnormal resistance noticed.Moreover, there were 3 other snares that were used on different patients that had similar issues that day.The procedure was completed with another captivator snare.There were no patient complications reported as a result of this event.The patient condition following the procedure was reported to be stable and fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIVATOR II
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
256 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key10593164
MDR Text Key208761759
Report Number3005099803-2020-04214
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729855903
UDI-Public08714729855903
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K133987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00561221
Device Catalogue Number6122
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-