Initial reporter city: (b)(6).Device evaluated by mfr.: returned product consisted of an epic self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the stent is partially deployed 3mm from the distal end of the middle sheath.Microscopic examination revealed no damages.The lock for the rack is missing.It was considered likely that the partial deploy was attributable to handling.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis did not find any kinks on the device.
|
Reportable based on device analysis completed on 21-sep-2020.It was reported that shaft kink occurred.The 99% stenosed target lesion was located in a severely tortuous and severely calcified shunt.An 8x100x75mm epic stent was selected for use.However, when the device was taken out, the shaft was found bent.The procedure was completed with a different device.There were no patient complications reported.However, returned device analysis revealed stent partial deployment.
|