The balloon catheter afapro28, with lot 72677-097 was returned and analyzed.External visual inspection showed the catheter was intact with no apparent issues.Verification of the smart chip file indicated that the catheter was used for fifteen injections on the date of the event.The catheter displayed a kink on the guide wire lumen under the balloon segment during the inflation test.The catheter passed the performed test.The dissection and pressure test did not show any leaks or blood inside the catheter.The test did show that there was a kink at 1.37 inches from the tip.In conclusion, the balloon catheter failed inspection due to the guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
|