Model Number 71360484
Device Problem Break (1069)
Patient Problem Foreign Body In Patient (2687)
Event Date 09/01/2020
Event Type  Injury  
Event Description
It was reported that during thr procedure the fem hd/neck impctr rpmt tip broke during impaction of the femur.No delay.Procedure could be finished with the same device.No other complications were reported at this time.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical / medical investigation concluded that, per complaint details the impactor ¿broke¿ during the femoral impaction of the tka procedure; however, the same device was reportedly used to conclude the procedure without delay.Responses to requests for medical documentation were not received.Patient impact beyond the reported events would not be anticipated as there was no patient injury alleged and no surgical delay reported.Therefore, no further medical assessment is warranted at this time.Should clinically relevant documentation/information become available, the clinical/medical task may be re-evaluated.A review of the complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEM HD/NECK IMPCTR RPMT TIP
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10597156
MDR Text Key208909792
Report Number1020279-2020-04918
Device Sequence Number1
Product Code HRY
UDI-Device Identifier00885556024362
UDI-Public00885556024362
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71360484
Device Catalogue Number71360484
Device Lot Number18KEM0001A
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/01/2020
Initial Date FDA Received09/28/2020
Supplement Dates Manufacturer Received08/05/2021
Supplement Dates FDA Received08/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-