Model Number WB991036
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hemorrhage/Bleeding (1888); Blood Loss (2597)
Event Date 09/22/2020
Event Type  Injury  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient¿s outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that after an unspecified therapeutic polypectomy procedure, bleeding occurred in the patient.The patient was then readmitted requiring a blood transfusion.No further information was provided.
 
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation.Therefore, the investigation/evaluation was performed exclusively on the basis of the information provided.According to the information provided,a periodic safety check (psc) was performed with the generator, where all tested parameters were found to be within specification.No defect, failure or malfunction could be determined.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected serial number of the hf generator without showing any abnormalities.Based on the information available, the exact cause of the reported phenomenon and the patient¿s outcome could not be conclusively determined.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OLYMPUS ESG-100, 220...240 V~,
Type of Device
HF-GENERATORS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key10598644
MDR Text Key208946855
Report Number9610773-2020-00224
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04042768910364
UDI-Public04042768910364
Combination Product (y/n)N
PMA/PMN Number
K073207
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 11/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberWB991036
Device Catalogue NumberWB991036
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/23/2020
Initial Date FDA Received09/29/2020
Supplement Dates Manufacturer Received11/24/2020
Supplement Dates FDA Received11/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-