The investigation determined that a lower than expected vitros amon result was obtained from a vitros liquid performance verifier (lpv) quality control fluid using vitros amon reagent on a vitros 5600 integrated system.A definitive assignable cause of the event was not established.A within run vitros amon precision was not processed, and it is unknown if any additional troubleshooting was performed aside from calibrating the vitros amon.Therefore, an analyzer issue cannot be confirmed or ruled out as a possible cause of the lower than expected vitros amon result.A vitros pv fluid issue is an unlikely contributor to the event, as acceptable pv results from both lpv i lot p8085 and lpv ii lot q8086 were obtained following the event.Fluid handling protocol was not provided therefore a quality control fluid handling issue cannot be ruled out as a contributor to the event.Quality control results showed unacceptable laboratory precision leading up to the day of the event but have been acceptable since the event.It is possible that day to day inconsistencies with lpv fluid handling is the cause of the unacceptable within laboratory precision, although this cannot be confirmed.Continual tracking and trending of complaint data has not identified any signals to suggest there is a systemic quality issue with vitros amon slides lot 1018-0253-7843.
|