Model Number 22716-4J |
Device Problems
Gas/Air Leak (2946); Material Twisted/Bent (2981)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
A review of the device history record is not possible as no lot number was provided.The actual complaint product was not returned for evaluation.Root cause could not be determined.All information reasonably known as of (b)(6) 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).
|
|
Event Description
|
It was reported that the sleeve inflated when the suction catheter was attached to a patient.Flapper valve was bent.The incident was found before suctioning and the suction catheter was replaced for new one.No patient injury was reported.Additional information received 10-sep-2020 indicated the product was not altered in any way prior to use.The patient has an endotracheal tube.No surgery or resuscitation was involved and there were no ventilator alarms.The tubing did not become unattached.There was no problem reported with the thumb valve and there was no need to extubate and reintubate the patient.
|
|
Manufacturer Narrative
|
Additional information: d4.A review of the device history record is in-progress.All information reasonably known as of 24 oct 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Manufacturer Narrative
|
The device history record for lot 30040413 was reviewed and the product was produced according to product specifications.The actual complaint product was not returned for evaluation.Photos of the device were provided and the reported incident was confirmed.Root cause was traced to manufacturing.All information reasonably known as of 04 dec 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|