U.S. Department of Health and Human Services

MAUDE Adverse Event Report: V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Failure to Charge (1085); Unintended Power Up (1162)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of events: (b)(6) 2020, date of report: (b)(6) 2020.
 
Event Description
The customer reported that the unit is turning on by itself and the battery is not charging.The customer contacted product support and requested for part id for user interface (ui) assembly and power management (pm) board.Product support provided the part id.The customer reported that the unit was not in use on a patient.
 
Manufacturer Narrative
G4:11jan2021.B4:12jan2021.The biomedical engineer replaced the power management (pm) and motor controller (mc) board to address the reported issue.No other fault was noted at time of service.The battery was replaced on jan 07, 2020.The unit is back in service.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key10603580
MDR Text Key209366094
Report Number2031642-2020-03508
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 09/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/10/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-