The device history records for the sam 300 device was reviewed and this confirmed that all manufacturing and quality checks and tests had been successfully completed.The sam 300 passed ¿out qat from heartsine technologies on the (b)(6) 2005.The device was returned with the reported fault of ¿memory full prompts¿.Further information from the reported complaint confirmed that the user had been regularly powering on the device to check the status of the battery.From device installation the device records a total of 63 log entries, which included 29 manual power cycles and 34 low battery fails.The large quantity of manual power ons in combination with the low battery fails had filled the device memory over the course of 11 years from installation, resulting in ¿memory full¿ prompts from the (b)(6) 2016.Additionally, the investigation found no fault with the device during testing at heartsine.The memory full prompt indicates the device is no longer able to record ecg/icg information but does not indicate a fault.The device can still analyse and deliver a shock if required, as demonstrated during the investigation.The returned sam 300 device is now out of warranty and will be scrapped.
|