Model Number 71369720
Device Problem Material Fragmentation (1261)
Patient Problem No Patient Involvement (2645)
Event Date 09/01/2020
Event Type  malfunction  
Event Description
It was reported that the mechanism which holds the head to the trial neck broke off, rendering the head unusable.As noticed in the mdrd, no case was involved.
 
Manufacturer Narrative
The device, intended use in treatment, was not returned for evaluation, the reported event could not be confirmed.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MI TRIAL FEM HEAD 36 -3
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10614185
MDR Text Key209411699
Report Number1020279-2020-05015
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010559913
UDI-Public03596010559913
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71369720
Device Catalogue Number71369720
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received12/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-