It was reported that this patient was admitted to the hospital intensive care unit (icu) due to cardiac arrest.The patient was noted to have a very low ejection fraction (ef).A review of data from the patients implanted cardiac resynchronization therapy defibrillator (crt-d) device indicated that the device had delivered twelve (12) total shocks across several ventricular tachycardia (vt) episodes on that day.The patients rhythm was noted to be polymorphic vt that progressed into fine ventricular fibrillation (vf).The shock therapy was noted to be appropriate but there were multiple shocks with unsuccessful arrhythmia conversion and in the first episode, therapy was exhausted.Boston scientific technical services (ts) indicated that following the first episode, the patients arrhythmia must have converted on its own as there were subsequent episodes with additional shock therapy provided and successful arrhythmia conversion.Ts noted however that there was questionable capture observed at the end of the second episode in addition to some undersensing observed of the fine vf signals.The presenting electrogram (egm) from that same day showed intrinsic atrial sensed beats and bi-ventricular pacing beats.The device remains in-service.No additional adverse patient effects were reported.
|